Temperature is the most common measurement on a test bench, and the first decision is the sensor type. Thermocouple vs RTD is a choice between range and robustness on one side and accuracy and stability on the other. This article compares the two sensor types, explains the wiring details behind most measurement errors, and lists the ICP DAS modules that read each type.
A thermocouple is a junction of two different metals. It produces a small voltage that depends on the temperature difference between the measuring junction and the point where the thermocouple wires meet copper, the cold junction. A type K thermocouple gives roughly 41 µV per °C, so the input must resolve microvolts and reject noise.
An RTD (resistance temperature detector) is a resistor whose value rises with temperature. A Pt100 has 100 Ω at 0 °C and changes by about 0.385 Ω per °C; a Pt1000 has 1000 Ω at 0 °C. The input module drives a small current through the sensor and measures the voltage across it.
| Property | Thermocouple | RTD (Pt100, Pt1000) |
|---|---|---|
| Temperature range | Wide: type T about −200 to +350 °C, types J and K up to about 750 °C and 1,200 °C, types R, S and B higher still | Typically −200 to +600 °C, some designs to +850 °C |
| Standard tolerance | Type K class 2: ±2.5 °C or 0.75 % of reading, whichever is larger | Class B: ±0.3 °C at 0 °C, ±0.8 °C at 100 °C |
| Long-term stability | Can drift, especially at high temperatures | Very stable |
| Signal | Microvolts; needs cold junction compensation | Resistance; needs excitation current and lead compensation |
| Response | Fast with thin or exposed junctions | Slower, depending on element and sheath size |
| Cable | Thermocouple or matching extension cable to the input | Copper cable, ideally 3 or 4 wires |
Choose an RTD when you need accuracy and stability between about −200 and +600 °C. Choose a thermocouple for higher temperatures, small or fast probes, and harsh mechanical conditions.
A thermocouple measures a difference. To get the absolute temperature, the module measures the temperature at its own terminals and adds the equivalent thermocouple voltage. This cold junction compensation (CJC) is only as good as the temperature match between the terminals and the CJC sensor, so mount modules away from heat sources and draughts. Run thermocouple or matching extension cable all the way to the input terminals; a copper section in between creates new junctions at an unknown temperature.
Among ICP DAS inputs, the ET-2218HS-16 and USB-4018HS compensate automatically on every channel. The M-7018-16-G and the SG-3011 conditioner provide an adjustable CJC offset for fine calibration. For PCI-1800U boards, the DB-8225 terminal board has a cold junction circuit on analog input channel 1.
In a 2-wire connection, the lead resistance adds directly to the sensor resistance.
Worked example: a Pt100 on 10 m of 0.5 mm² copper cable sees about 0.34 Ω per conductor, or 0.69 Ω in the loop. Divided by 0.385 Ω/°C, the reading is about 1.8 °C too high. The same cable with a Pt1000 gives an error of about 0.18 °C, because the sensor changes ten times as much per degree.
A 3-wire connection uses a third lead to measure and subtract the lead resistance, assuming all leads are equal, and is the usual choice for test-bench wiring. A 4-wire connection carries the excitation current on one pair and measures the voltage on the other, which removes lead resistance from the result; use it for calibration and reference measurements.
The M-7015P-G, ET-2215H, ET-7015 and ET-7215 provide 3-wire lead resistance elimination. The M-7033-G accepts 2-, 3- and 4-wire RTDs, with lead resistance elimination for 3- and 4-wire connections.
Sensor response time depends mostly on construction and thermal contact with the test object. Thin, exposed thermocouple junctions react fastest; a sheathed probe of either type reacts more slowly as its diameter grows.
The input module must also sample fast enough. Standard multi-channel modules share their sample rate across channels: the M-7018-G samples 10 Hz in total over 8 channels and the M-7015-G 12 Hz over 6, which suits slow thermal tests. For fast transients, the ET-2218HS-16 and USB-4018HS sample thermocouples at 100 Hz per channel, and the ET-2215H samples RTDs at 90 Hz per channel in fast mode, with ±0.1 % of FSR accuracy against ±0.05 % in normal mode at 1.5 Hz. The SG-3011H thermocouple conditioner has a switch-selectable response time of 0.5 ms or 100 ms.
For DAQ boards or PLCs with voltage or current inputs, the SG-3011 (thermocouple) and SG-3013 (RTD) convert one sensor to a voltage or current output, with 1000 VDC three-way isolation.
Send your sensor types, temperature ranges, channel count and required sample rate to our technical sales team, and we will suggest suitable modules.
The PET-7017-10 is a PoE-powered Ethernet analog input module for engineers who need voltage or current measurement on a networked I/O node. Its inputs can be set as 10 differential or 20 single-ended channels in software, and each channel can be given its own range, so one module can serve mixed sensor types on the same network.Each channel offers a choice of voltage ranges from ±150 mV to ±10 V and current ranges of ±20 mA, 0 ~ 20 mA and 4 ~ 20 mA, selected with jumpers. Resolution is 16-bit, and the module has two sampling modes: normal mode at 10 Hz total and fast mode at 50 Hz total. Open wire detection is available for 4 ~ 20 mA inputs, and differential channels are rated for 240 Vrms overvoltage protection.Configuration and monitoring are handled through a built-in web server with Web HMI, and the module supports Modbus TCP and Modbus UDP. Access is controlled by ID, password and IP filter. The unit is powered either from PoE (IEEE 802.3af, Class 1) or from a +12 ~ +48 VDC terminal block, and it mounts on a DIN rail.ApplicationsBuilding automationFactory automationMachine automationRemote maintenanceRemote diagnosisTesting equipment
DN-AIH-08 is a termination board for HART analog input modules. It measures the analogue signal on each of its 8 channels and can also supply isolated loop power to field devices, so it accommodates both active and passive loop wiring.The board includes short-circuit protection on the isolated power output. If the circuit is shorted or the current becomes too high, the loop current is limited. Removable fuses protect loop devices from high inrush current, and each fuse has a physical indicator that shows when it has blown.Suited to HART analog input installations where field devices need to be wired in either active or passive loop configuration.Mounts on DIN-rail or wall, with dimensions of 135 x 214.6 x 25.5 mm.ApplicationsConnecting HART analogue field devices to HART analog input modulesLoop-powered field instrumentation requiring isolated passive loop powerIndustrial installations requiring protected analogue loop wiring
DN-AII-08 CR is a termination board for current analog input modules. It connects field signals to the input channels and also measures the analog current, so it fits systems where the input is a current loop.The board supplies isolated loop power to field devices and supports both active and passive loop wiring on its 8 channels. The isolated supply is short-circuit protected: the loop current is limited when the circuit is shorted or the current becomes excessive. Removable fuses protect loop devices from high inrush current, and each fuse shows a physical indicator when it has blown.The board measures 135 x 214.6 x 25.5 mm and mounts on a DIN rail or on a wall. It is rated for an operating temperature of -25 ~ 75 °C and 5 ~ 95% RH, non-condensing.ApplicationsCurrent-loop field device wiring with active or passive loops
The DN-AIO-M CR is a universal termination board for analog input and output modules. It is designed as a general-purpose wiring point for most kinds of analog module, so one board can serve a range of analog I/O installations.Its footprint is small, matching other mini termination boards, which makes it easier to fit into control panels where space is limited. It can be mounted on a DIN rail or on a wall.Pin assignment and wiring information covers voltage and current input and output configurations.Operating range of -25 ~ 75 °C and 5 ~ 95% RH, non-condensing.Carries CE, UKCA, FCC, RoHS and WEEE markings.ApplicationsTerminating field wiring for analog input modulesTerminating field wiring for analog output modulesCompact wiring points in control panels and installations
How IEPE accelerometers work, how to set excitation, gain and sample rate, and which ICP DAS conditioners and DAQ modules suit vibration tests.
Sample rate, simultaneous sampling, latency, isolation, distance and drivers compared for PCI/PCIe DAQ boards, USB I/O and Ethernet I/O.
Vi kjenner ICP DAS-sortimentet og hjelper deg med å finne riktig modul, tilbehør eller reservedel.
Kontakt teknisk salg